Remote Control
The BeamView Analyzer provides remote control and
data transfer through the host computer’s RS-232 (and
TCP/IP with BeamView-USB). A complete control and data
transfer command set is provided to allow users to develop
their own remote control application for interfacing with
the BeamView Analyzer software platform. The BeamView-
USB software package includes a full suite of LabVIEW
libraries for remote access to most BeamView features at
a host computer running LabVIEW.
Beam Analysis and Statistics
BeamView Analyzer software calculations are compatible
with the International Standards Organization (ISO)
guidelines for laser beam measurement:
Peak and centroid beam position
Beam ellipticity including angular position and
major/minor axis information
Circularity
D4σ diameters and widths
Guassian t including coefcient, centroid, and “roughness of t”
Continuous on-line statistical analysis display
Calculations Pass/Fail test settings
Polar beam wander plot screen Fault Actions
Dialog Box
Beam Stability
The continuous on-line statistical analysis display shows
results of all, or a combination of, functions and pass/fail
parameters for all captured samples and accumulated
results. The user can scroll through the analysis results of
individual images, and also view the minimum, maximum
and sigma (standard deviation) values. This makes
comparing individual samples to the time-dependent
statistical data easy. Thus, the jitter and stability of
parameters, such as power, energy, pointing direction,
ellipticity and beam size, etc., can be analyzed
simultaneously with a polar beam wander plot.
Pass/Fail Analysis
Pass/fail analysis allows simultaneous real-time monitoring
of all, or any one of the analysis results against user-specied
minimum/maximum limits. Any combination of, or all the fault
actions can be activated to signal a test failure, initiate a visual
alarm, an audio alarm, stop data capture, reject/save a failed
sample, and generation of a TTL
trigger pulse output signal.
Laser
Cross-
Reference
Index
POWER
& ENERGY
Power
Sensors
Energy
Sensors
Custom
& OEM
BEAM
DIAGNOSTICS
CALIBRATION
& SERVICE
INDICES
Product
Name
Index
Power
& Energy
Meters
BeamView Analyzer Software Features
BeamView Analyzer Software
Aperture t and uniformity
Total/relative power
Peak power/energy density
Percent power within an aperture
The BeamView analyzer system also includes several other analysis features listed in the BeamView Analyzer Software Features