44
Measurement
Display and
Range
Selection
The display update rate for numeric measurement is
three times per second.
Measurement range is selected in decade steps. Range
selection (shown in Table 3 on page 45, and Table 4 on
page 46) is dependent on the sensor type and character-
istics, as well as user-determined measurement settings.
FieldMaxII-TOP uses the “3’s” Rule—a display format-
ting rule in which the display value is not allowed to
exceed 3, 30, or 300, depending on where the decimal
point falls with the decimal point located in a fixed posi-
tion, as determined by the current range. Typically if a
reading exceeds the “3’s” limit, the instrument is over-
ranged or, in the case of Auto Ranging, the instrument
will automatically range up.